User Entry of the Agilent (HP) 8510 Calibration Standards Definitions for Millimeter Wave Configuration
The purpose of this procedure is to guide a 8510
millimeter wave user through the completion of the OML Calibration Standards
Definition Chart (OML Definitions Chart).
The procedure will then address the Agilent 8510 (8510) front panel
inputs necessary for entry of the characteristics of Waveguide Standards made up
of the components of an OML millimeter wave waveguide Calibration Kit.
When the Calibration Standards Definitions have been entered, the 8510
user will then be able to perform a waveguide calibration procedure.
Calibrations, based on these Calibration Standards Definitions, are
possible for both the "one
path-two port" measurement utilizing one OML T/R type module, one OML T
type module and the "two path-two port" utilizing two OML T/R type
modules. Carefully follow the
procedure on a step-by-step basis for maximum user efficiency.
This procedure assumes that the users 8510 system has been correctly
configured and is operating properly.
For reference, the configuration process is described in the OML paper
entitled User Control of the HP 8510 for Millimeter Wave
Configuration or in the OML millimeter wave module user's manual.
This procedure follows closely the Agilent
methodology for the calibration process.
A calibration kit is composed of several components which are
individually or collectively referred to as "Calibration Standards."
Agilent has assigned an arbitrary "class" number and a
definition methodology to each of these Calibration Standards.
The mix of Standards to be used is determined by the calibration method
to be performed. A listing of
the various type of calibrations typically used for millimeter wave waveguide
applications is found in the OML document Most Common
Types of Calibration Procedures or in the OML millimeter wave module user's
manual. Included in that
document are the types of "Calibration Standards" used in each
calibration procedure.
Millimeter
Wave Waveguide VNA Calibration Procedures vs. OML Cal Kit Components lists
the various calibration procedures that can be performed with an OML Cal Kit.
An
example would be the SSLsT calibration method which is popular as millimeter
wave waveguide calibration. It
utilizes the following "Calibration Standards": 1) a Short, 3) an
Offset Short, 10) a Load (sliding) and 11) a Thru connection.
The number, x), shown in front of each item is the Class assignment for
that Standard as defined by Agilent. OML
has chosen to use the same "class assignments" and definition
procedure as Agilent with a few minor deviations or amplifications to simplify
the process. The OML
Definition Chart that follows combines the functions of both the Agilent
"Standard Class Assignments Chart" and the Agilent "Calibration
Standards Definition Chart." This
is possible by the use of the same Standards Class assignments reflected in
Agilent's calibration kits as being applicable to the Class assignments for the
Standards in the OML calibration kits.
Calibration kits are available from several vendors
covering the waveguide bands up through 110 GHz.
One of the vendors also has a calibration kit for WR-08 covering up to
140 GHz. OML is the only
vendor offering calibration kits for all waveguide bands covering from 33 to 325
GHz. Almost all Cal Kits
include the following components: 1)
a flush short, 2) a ¼ wavelength
section, and 3) a fixed load. Some
kits include a precision waveguide section, of known length, to be used as a
verification tool for the quality of the calibration, and one or two waveguide
sections to be used as "test port adapters" which are to be attached
to the waveguide ports of the millimeter wave VNA modules.
Typically, all vendors supply these components with precision waveguide
flanges. The OML Millimeter
Wave VNA Cal Kits are similar to those provided by other vendors but differ in
that they includes all of the above components plus:
an additional flush short, an additional waveguide load and a sliding
waveguide load. The OML Cal
Kit includes a precision waveguide section, of known length, to be used as a
verification tool for the quality of the calibration.
The OML Cal Kit provides two precision waveguide shims whose difference
is ¼ wavelength instead of a single ¼ wavelength shim.
The first shim is arbitrarily 0.1000" thick and the second shim is
0.1000" plus l/4
at the geometric mean frequency (GM(Fo)) of the waveguide band being addressed.
This results in shims that are more robust (resistant to damage) and are
easier to produce accurately up to 325 GHz.
All of the components in the
OML Cal Kit are produced using electro-forming
techniques offering the best in dimensional accuracy and are mechanically
traceable to N.I.S.T.
To illustrate the "Calibration Standards
Definitions" process let us look at typical data and walk through the
necessary steps. The OML Cal
Kit includes eight components which used either individually, or in combination
with another component, compose a Standard.
Included with each OML Calibration Kit is a chart (see pdf flie "Table
1.
Example "OML VNA
Calibration Kit" data sheet) listing the pertinent properties of each of
the Standard components. The
accuracy of the mechanical measurement of dimensional properties is traceable to
N.I.S.T. The "delay"
of each of the Standards is calculated from the length dimension.
These components include (the example shown is for a WR-12
60 to 90 GHz Cal Kit): a) a
waveguide termination (labeled "T-12 Precision Term"), b) a second
waveguide termination (also labeled "T-12 Precision Term"), c) a flush
short (labeled "S-12 Precision Short"), d) a second flush short (also
labeled "S-12 Precision Short"), e) a 0.1000" waveguide shim
(labeled "OS-12 Precision Insert B"), f) a 0.1000" plus l/4 waveguide shim (labeled "OS1/4-12
Precision Insert A"), g) an adjustable waveguide load (labeled "AL-12
Adjustable Load") and h) a waveguide section of precisely known length
(labeled "W/G-12 Precision Section").
An example situation: a "one path-two port" calibration of a
WR-12 system using an OML V12VNA-T/R module (T/R), with its test port adapter
waveguide section attached, and an OML V12VNA-T module (T), with the supplied
waveguide attenuator attached. An
SSL(s)T calibration would make use of either the first or second flush short (no
difference exists) in combination with the 0.1000" waveguide shim to form
the Standard for the Short, the first "S" in SSL(s)T.
Either the first or second flush short in combination with the
0.1000" plus l/4 waveguide shim would be used to form the
Standard for the Offset Short, the second "S" in SSL(s)T.
The sliding load will be used as the L(s) Standard for the SSL(s)T.
A connection of the T/R module test port (with the Test Port Adapter
attached) to the test port of the T module input of the waveguide attenuator
(attached to the input of the T module form the Thru, the T in SSL(s)T.
OML Calibration Standards Definitions Chart use
The OML Definitions Chart was intentionally made
similar to the Agilent version so as to allow the user ease of transition back
and forth between the OML and Agilent versions.
As the OML Definitions Chart is concerned only with millimeter wave
waveguide calibration, all of the items on the Agilent Definitions Chart
relevant to coaxial calibrations were excluded.
Again the Agilent "Class" assignments used for their
calibration kits were adopted by OML.
For the discussion of the OML Definitions Chart the
WR-12 Cal Kit example above (pdf file Table 1) will be used.
The pdf file Table 2 is an example of an OML Definitions Chart with the data from a
typical OML WR-12 Cal Kit inserted. Starting
from the left of the OML Definitions Chart the first column is divided into two
sub-columns. The "Class
#" column is used to determine the Class # assignment and replaces the
Agilent "Standard Class Assignment Chart" for our purposes.
The "Type" column is used to identify which of the three
waveguide calibration standard type is assigned to that Class #.
The three type of Standards are Short, Load and Delay/Thru.
The term Short can represent flush short, an offset short or a sliding
short (typically not used in waveguide calibration). The
term Load can represent a fixed load, an offset load and a sliding load.
The term Delay/Thru recognizes that a through connection will exhibit
some delay property. The
direct interface of the test ports of two millimeter wave modules will have be a
Thru connection of zero delay. The
waveguide shims or the "known length" waveguide section will each have
a different delay property and can be used as part of the "Thru"
calibration. The choice of
which "Type" of standard is assigned to which "Class #" is
arbitrary. However, for a
given calibration kit, once the data is entered into the 8510, the "Class
#" of a standard "Type" cannot be changed.
There are two "Offset" columns, one for the
waveguide length (delay) data of a particular Standard, and one to indicate that
the is system mode is waveguide. The
"Delay pS" of a Standard is derived from it's insertion length. ).
Side note, the delay data must be entered into the 8510 in "nS"
but all of the Agilent charts and the 8510 display will show the delay in "pS"
for the waveguide bands covered by OML.
Each of the Standards will have a length entry ranging from "0"
(no length) to that of the longest Standard to be used.
The entry for Standard 1, representing the Short of a Short/Offset Short
pair, will be that of the combination of a OML "S-12" flush short and
the OML "OS-12" 0.1000" waveguide shim which will have a delay of
0.008475 pS (the delay of the 0.1000" waveguide shim only).
Likewise, the entry for Standard 3, representing the Offset Short of a
Short/Offset Short pair, will be that of the combination of a OML
"S-12" flush short and the OML "OS-1/4-12"
0.1000" plus l/4 waveguide shim which will have a delay of
0.012993 pS (the delay of the 0.1000" plus l/4
waveguide shim only). The
Offset column labeled "Zo W"
is concerned with the 8510 front panel entry used to denote the system test port
impedance, i.e., coax or waveguide mode.
A value of "1" should be entered for each Standard indicating
that the system is in the waveguide mode.
There are two entries necessary for "Frequency
(GHz)". The first is for
the TE 1,0 cutoff frequency and is labeled "Min".
The TE 1,0 cutoff frequency, 48.362 (GHz), of the WR-12 waveguide is
indicated in our sample chart. The
second is for the TE 2,0 cutoff frequency and is label "Max".
This is also known as the "mode" frequency and is twice the TE
1,0 frequency. The entry on
our sample chart is 96.724 (GHz). These
two entries must be a very precise entries as the VNA uses this data for a
number of it's algorithms in the calibration process.
This is especially true as the user goes higher in frequency above 110
GHz (WR-10) as waveguide vendors do not guarantee their waveguide dimensions for
WR-08 and above . The use of
the commonly published waveguide cutoff frequencies can cause potentially fatal
errors in the calibration. Use
the OML Cal Kit data chart TE 1,0 and TE 2,0 frequency data for the precision
components in the OML Cal Kit. For
more information see the OML paper "Millimeter Wave Vector Analysis
Calibration and Measurement Problems Caused by Common Waveguide
Irregularities" that is on the OML web site or in the OML VNA modules users
manual.
The formal description of the Standard is entered in
the column "Standard Label".
This is also description that is entered into the 8510.
In the following list the Standard
Labels is shown along with the Standard Class #, The Standard Type and the
OML Cal Kit Components used to assemble each Standard.
The Standards possible with the components of an OML WR-12 Cal Kit are
used as examples.
Standard Class Standard
Type
Standard Label
OML Components
1
Short
SHORT
S-12 + OS-12
3
Short
/OFFS
S-12 + OS1/4-12
9
Load
FIXED
T-12
10
Load
SLIDING
AL-12
11
Delay/Thru
THRU
none
13
Delay/Thru
THRU W/G
W/G-12
14
Delay/Thru
THRU .1
OS-12
15
Delay/Thru
THRU.1+/4
OS-12
20
Load
OFFS LOAD
T-12 + OS-12
21
Load
/4OFFS LD
T-12 + OS1/4-12
OML has made use of the latitude in labeling that is
available in the 8510 "LABEL STD" function to make the labels more
meaningful. Standards 13
through 21 are typically not possible with other vendor's Cal Kits.
An OML paper entitled "Millimeter Wave Waveguide VNA Calibration
Procedures vs OMLCal Kit Components" can be found in the OML millimeter
wave module user's manual. This
paper details which of the OML Cal Kit components are used as, or used to make
up the Standards required for the various calibration procedures.
The next two columns are used by the user for
"record keeping" and are not for data to be entered into the 8510.
The OML Cal Kit components that make up each Standard are identified in
the column labeled " OML Components".
For WR-12 the Standard 1
consists of the combination of an OML "S-12" short and the OML
"OS-12" 0.1000"
waveguide shim. The
"Comments" column is provided for notation of a verbal description of
the Cal Kit components or combination thereof used for a given Standard or for
specific calibration conditions to be utilized, etc.
Entering the Calibration Standards Definitions into
the 8510
The proper operation of the millimeter wave 8510 VNA
system should confirmed prior to the beginning of any calibration effort.
The OML document "User Control of the HP 8510 for Millimeter Wave
Configuration" should be followed precisely in setting up the 8510 system.
Upon completion of "set-up" the user should examine the output
signal of each downconverter channel.
Using the "MENU" under "PARAMETERS" the module
Reference signal output can be viewed by pressing the "A1" or
"A2" buttons depending on the system configuration.
Likewise, the module Test signal output can be viewed by pressing
The Cal Kit data that has been entered by the user on
to the "Calibrations Standards Definitions" chart supplied in the OML
Millimeter Wave VNA Module Users Manual (the data found on the Calibration Kit
data sheet) is to be entered into the Agilent 8510 using the following steps.
1) Press the “
2) Press the “MORE” soft
key
3) Press the “MODIFY
1" soft key (MODIFY 2 can be used if preferred).
The data for the following steps is that associated
with Standard 1, the "OS-xx Precision Insert B" offset shim
(0.1000") component of the OML Calibration.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "1"
key on the "ENTRY" Keypad followed by “X1".
3) Press the “SHORT” soft
key
At this point entries for Short properties
"L0" through "L3" appear adjacent to soft keys.
These are coaxial properties and do not apply to waveguide
"Short" Standards. Simply
make sure that "0" in each of these entries.
If the entry is other than "0" perform the following steps.
4) Press the "L0"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
5) Press the "L1"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
6) Press the "L2"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
7) Press the "L3"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
8) Press the “SPECIFY
OFFSET” soft key
9) Press the “OFFSET
DELAY” soft key
10) Convert the data listed
in picoseconds for Standard 1 (OS-xx Precision Insert B) on the
"Calibrations Standards Definitions" chart to nanoseconds and
enter that number followed by
"G/n" using the "ENTRY" keypad.
Example: ".008475"
and "G/n" (8.475 pS) for a typical
OML WR-12 Cal Kit.
11) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
12) Press the “OFFSET Zo W” soft key and enter "1" followed
by "x1" using the "ENTRY" keypad.
13) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
14) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
15) Press the
"WAVEGUIDE" soft key.
16) Press the "STD
OFFSET DONE" soft key.
17) Press the LABEL STD"
soft key.
18) If the label shown is
"SHORT" proceed, if not enter "SHORT".
19) Press the "TITLE
DONE" soft key.
20) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 3, the "OS1/4-xx Precision Insert A" offset shim
(0.1000" plus 1/l)
component of the OML Calibration Kit.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "3"
key on the "ENTRY" Keypad followed by “X1".
3) Press the “SHORT” soft
key
At this point entries for Short properties
"L0" through "L3" appear adjacent to soft keys.
These are coaxial properties and do not apply to waveguide "Short
Standards. Simply make sure
that "0" in in each of these entries.
If the entry is other than "0" perform the following steps.
4) Press the "L0"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
5) Press the "L1"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
6) Press the "L2"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
7) Press the "L3"
soft key and enter "0" using the "ENTRY" keypad followed by
"x1".
8) Press the “SPECIFY
OFFSET” soft key
9) Press the “OFFSET
DELAY” soft key.
10) Convert the data listed
in picoseconds for Standard 1 (OS1/4-xx Precision Insert A) on the
"Calibrations Standards Definitions" chart to nanoseconds and
enter that number followed by
"G/n" using the "ENTRY" keypad.
Example: ".012993" & "G/n" (12.993 pS) for the
sample.
11) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
12) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
13) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
14) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
15) Press the
"WAVEGUIDE" soft key.
16) Press the "STD
OFFSET DONE" soft key.
17) Press the LABEL STD"
soft key.
18) If the label shown is
"/4 OFFS" proceed, if not enter "/4 OFFS".
19) Press the "TITLE
DONE" soft key.
20) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 9, one of the "T-xx Precision Term" waveguide load
components of the OML Calibration.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "9"
key on the "ENTRY" Keypad followed by “X1".
3) Press the “LOAD” soft
key
4) Press the
"FIXED" soft key
5) Press the “SPECIFY
OFFSET” soft key
6) Press the “OFFSET
DELAY” soft key and enter "0" followed by "x1" using the
"ENTRY" keypad.
7) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
8) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
9) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
10) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
11) Press the
"WAVEGUIDE" soft key.
12) Press the "STD
OFFSET DONE" soft key.
13) Press the LABEL STD"
soft key.
14) If the label shown is
"FIXED" proceed, if not enter "FIXED".
15) Press the "TITLE
DONE" soft key.
16) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 10, the "AL-xx Adjustable Load" sliding waveguide load
component of the OML Calibration Kit.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "10"
key on the "ENTRY" Keypad followed by “X1".
3) Press the “LOAD” soft
key
4) Press the
"SLIDING" soft key
5) Press the “SPECIFY
OFFSET” soft key
6) Press the “OFFSET
DELAY” soft key and enter "0" followed by "x1" using the
"ENTRY" keypad.
7) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
8) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
9) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
10) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
11) Press the
"WAVEGUIDE" soft key.
12) Press the "STD
OFFSET DONE" soft key.
13) Press the LABEL STD"
soft key.
14) If the label shown is
"SLIDING" proceed, if not enter "SLIDING".
15) Press the "TITLE
DONE" soft key.
16) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 11, a "DELAY/THRU".
This is a zero delay, direct connection of two OML Millimeter Wave VNA
Modules with the Test Port Adapters (the input attenuator for a "T"
module) in place and no other devices inserted.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "11"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "THRU"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter "0" followed by "x1" using the
"ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) If the label shown is
"THRU" proceed, if not enter "THRU".
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The following entries, for Standards 13 through 21,
are optional. These describe
advanced Standards that it is possible to configure using the components of the
OML Cal Kit. The user should
consult the OML document "Most Common Types of Calibration Procedures"
available at http//www.omlinc.com/vna_cal.htm
or in the OML millimeter wave module user's manual.
Additional information can be found in the other Agilent documents
referenced at the end of this document.
The question of does the application require these additional calibration
delays should be considered before undertaking calibrations using these
Standards.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "13"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "THRU"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter delay of the precision waveguide section
(converted
to nS) followed by "nS" using the "ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) Enter "THRU
W/G" using the wheel and soft keys.
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 14, a "DELAY/THRU".
This is the 0.1000 waveguide shim inserted between two OML Millimeter
Wave VNA Modules with the Test Port Adapters (the input attenuator for a
"T" module) in place.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "14"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "THRU"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter delay of the 0.1000" waveguide shim
(converted
to nS) followed by "nS" using the "ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) Enter "THRU .1"
using the wheel and soft keys.
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 15, a "DELAY/THRU".
This is the OS-xx 0.1000 + l/4 waveguide shim inserted between two OML
Millimeter Wave VNA Modules with the Test Port Adapters (the input attenuator
for a "T" module) in place.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "15"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "THRU"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter delay of the 0.1000" + l/4
waveguide shim
(converted
to nS) followed by "nS" using the "ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) Enter "THRU.1+/4" using the wheel and soft keys.
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 20, a "LOAD".
This is the OS-xx 0.1000"
waveguide shim inserted between a T-xx Precision Term" waveguide load and
the Test Port Adapter attached to an OML Millimeter Wave VNA T/R Module.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "20"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "LOAD"
soft key
3a) Press the "OFFSET"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter delay of the 0.1000" waveguide shim
(converted
to nS) followed by "nS" using the "ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) Enter "OFFS
LOAD" using the wheel and soft keys.
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The data for the following steps is that associated
with Standard 21, a "LOAD".
This is the OS1/4-xx 0.1000 + l/4
waveguide shim inserted between a T-xx Precision Term" waveguide load and
the Test Port Adapter attached to an OML Millimeter Wave VNA T/R Module.
1) Press the “DEFINE
STANDARD" soft key
2) Press the "21"
key on the "ENTRY" Keypad followed by “X1".
3) Press the "LOAD"
soft key
3a) Press the "OFFSET"
soft key
4) Press the “SPECIFY
OFFSET” soft key
5) Press the “OFFSET
DELAY” soft key and enter delay of the 0.1000" + l/4
waveguide shim
(converted
to nS) followed by "nS" using the "ENTRY" keypad.
6) Ignore the “OFFSET
LOSS” soft key, it is not relevant as the entries are for the waveguide mode.
7) Press the “OFFSET Z0”
soft key and enter "1" followed by "x1" using the
"ENTRY" keypad.
8) Press the "MINIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using
the "ENTRY" keypad.
9) Press the "MAXIMUM
FREQUENCY" soft key and enter that frequency followed by "G/n"
using the "ENTRY" keypad.
10) Press the
"WAVEGUIDE" soft key.
11) Press the "STD
OFFSET DONE" soft key.
12) Press the LABEL STD"
soft key.
13) Enter "/4OFFS LD" using the wheel and soft keys.
14) Press the "TITLE
DONE" soft key.
15) Press the "STD
DONE" soft key.
The next step will be to enter the Calibration
Standards Definitions data into the 8510.
The OML procedure "XXXXX" will detail this process.
The following documents will aid the user in implementing this procedure
and for developing greater insight into 8510 calibration:
1) HP Product Note 8510-5B (5C coming soon),
Network Analysis, “Specifying calibration standards for the HP 8510
network analyzer”, 2) HP Product
Note 8510-8A, Network Analysis,
“Applying the HP 8510 TRL calibration for non-coaxial measurements”,
3) HP Product Note 8510-12. Millimeter-Wave
Measurements, “Using the HP 8510 Network Analyzer”,
4) HP 11644A Series Waveguide Calibration Kits Operating and Service
Manual, 5) The flow charts for the
calibration procedures found in the 8510C Operating Manual.
Updated 11/01/2002